The federal regulation
governing laboratory testing, known as the Clinical
Laboratory Improvement Amendments of 1988 (CLIA '88),
classifies testing into four complexity levels: high
complexity, moderate complexity, PPM (Provider Performed
Microscopy, a sub-set of moderate complexity), and waived
testing. |
Effective January 1, 2005
Standard PC.16.1022
The hospital
establishes policies and procedures that define the context
in which waived test results are used in patient care,
treatment, and services.
Elements of
Performance for PC.16.1022
-
Quantitative test result
reports in the clinical record are accompanied by
reference intervals specific to the test method used and
are appropriate to the population served.
-
Criteria for confirmatory
testing for each test, qualitative or quantitative, is
specified in the written procedure as dictated by clinical
usage and methodology limitations.
-
Actual usage is
consistent with the hospital's policies and the
manufacturer's recommendations for each waived test.
Standard PC.16.20
The hospital identifies the staff responsible for performing
and supervising waived testing.
Elements of
Performance for PC.16.20
-
Staff members who perform
testing are identified.
-
Staff members who direct
or supervise testing are identified.
Standard PC.16.30
Staff performing tests have adequate, specific training and
orientation to perform the tests and demonstrate
satisfactory levels of competence.
Rationale for PC.16.30
For waived
tests to be performed properly, the staff performing them
must be qualified to do so. Staff members who perform waived
testing have specific training in each test performed.
Elements of
Performance for PC.16.30
-
Current competence of
testing staff is demonstrated.
-
Each staff member who
performs testing has been trained specifically to each
test he or she is authorized to perform.
-
Each staff member who
performs testing has been oriented according to the
hospital's specific needs.
-
Testing that requires the
use of an instrument is performed by staff with adequate
and specific training on the use and care of that
instrument.
-
Competence is assessed
according to hospital policy at defined intervals, but at
least at the time of orientation and annually
thereafter.
-
These assessments have
considered the following:
-
The frequency by which
staff members perform tests; the technical backgrounds
of the staff.
-
The complexity of the
test methodology and the consequences of an inaccurate
result
-
Methods to assess current
competency include at least two of the following:
-
Performing a test on an
unknown specimen
-
Having the supervisor
or qualified delegate periodically observe routine work
-
Monitoring each user's
quality control performance
-
Having written testing
that is specific to the method assessed
-
The hospital evaluates
and documents the information listed above.
Note:
All staff who perform instrument-based
testing, including but not limited to physicians,
licensed independent practitioners, contracted staff, and
RNs, must participate in training and competence
demonstrations.
Standard PC.16.40
Approved policies and procedures governing specific
testing-related processes are current and readily available.
Rationale for PC.16.40
Current and
up-to-date policies and procedures are an important
reference tool in managing laboratory testing activities,
particularly when individual staff members perform them
infrequently.
Elements of
Performance for PC.16.40
-
Written policies and
procedures address specific sections<<>>
-
The policies and
procedures for each item are applicable to the specific
hospital.
-
Current and complete
policies and procedures are readily available to the
person performing the test.
-
The director named on the
waived testing certificate or a designee approves policies
and procedures at defined intervals.
Standard PC.16.50
Quality
control checks, as defined by the hospital, are conducted on
each procedure.
Elements of
Performance for PC.16.50
-
The hospital has a
written quality control plan that specifies how procedures
will be controlled for quality, establishes timetables,
and explains the rationale for choice of procedures and
timetables.
-
Quality control
procedures are performed at least as frequently as
recommended by the manufacturer, according to the
hospital's policies.
-
For instrument-based
waived testing, quality control requirements include two
levels of control, if commercially available.
-
Quality control
procedures are performed at least once each day on each
instrument used for patient testing.
-
The documented quality
control rationale is based on the following:.
-
How the test is used
-
Reagent stability
-
Manufacturers'
recommendations
-
The hospital's
experience with the test
-
Currently accepted
guidelines
-
At a minimum,
manufacturers' instructions are followed.
Standard PC.16.60
Appropriate quality control and test records are maintained.
Elements of
Performance for PC.16.60
-
All quality control test
results are documented, including internal, external,
liquid, and electronic.
-
Test results are
documented. Note: Test results may be located in
the clinical record.
-
Quality control records,
instrument problems, and individual results are
correlated.
-
A
formal log is not required, but a functional audit trail
is maintained that allows retrieval of results and
associated quality control values for a minimum of two
years.
|